The major hardware components in a mass spectrometer are:


Mass Spectrometer
After a molecule is introduced into the mass spectrometer, the molecule is first ionized, then fragmented. Following this, the ions can be selected and counted. A plot of the ion abundance versus the mass-to-charge ratio is a mass spectrum.
Play the video Download the video (0.7 Mbytes) Return to top of page

Vacuum System
For proper operation, the interior of the mass spectrometer must be evacuated. The ion source, mass filter, and detector all are under vacuum. The vacuum system makes it possible for ions to move from the ion source to the detector without colliding with other ions and molecules.

Achieving a vacuum sufficient for effective mass spectrometer operation generally requires two pumps. A high vacuum is achieved by connecting a low-vacuum pump to the output of a high-vacuum pump.

Two frequently used pump combinations are:

  • Diffusion pump(s) + rotary-vane rough pump
  • Turbomolecular pump(s) + rotary-vane rough pump
    Play the video Download the video (0.7 Mbytes) Return to top of page
    Mean Free Path
    For a charged particle to behave as directed by the electric and magnetic fields, it must not be subject to forces from collisions with other molecules. Therefore, mass spectrometry requires a vacuum system such that, on average, a charged particle will undergo no more than a few low-energy collision between the time of ionization and detection. The mean-free path of a charged particle is usually greater than the distance between ionization and detection.

    Pressure (Torr)
    Mean Free Path (meters)
    760
    6.0x10-8
    1
    4.5x10-5
    10-3
    4.5x10-2
    10-5
    4.5x10(0)
    10-7
    4.5x102
    10-9
    4.5x104


    Sample Inlet
    The first step to producing a mass spectrum is directing molecules into the mass spectrometer through the sample inlet. Typically, the molecules go from atmospheric pressure to a lower pressure as they move through the inlet.

    Gas, liquid, and solid compounds can be introduced into the ion source through specially designed inlets. This can be done a number of different ways depending on the state of the sample and the capabilities of the mass spectrometer vacuum system.

    Several of the common types of sample inlets are:

  • Direct and split inlets for use with GCs
  • Direct Insertion Probe (DIP)
  • Batch

    In all cases, the sample must be in the vapor state when it enters the ion source.

    Play the video Download the video (0.3 Mbytes) Return to top of page

    Ion Source
    Sample molecules are introduced into the ion source through the sample inlet. Before the mass spectrometer can analyze a sample, the sample molecule must be ionized.

    In the ion source, the molecules go through an ionization and fragmentation process. The ions are then directed into the mass filter.

    In electron ionization (EI), relatively high-energy electrons (70 eV) collide with molecules of the sample that is analyzed. These collisions produce (primarily) positive ions. Upon ionization, the molecules of a given substance fragment in fairly predictable patterns. EI is a direct process: energy is transferred collisionally from electrons to the sample molecules. <

    Play the video Download the video (0.7 Mbytes) Return to top of page
    For chemical ionization (CI), in addition to the sample and carrier gas, large amounts of reagent gas are introduced into the ionization chamber. Because there is so much more reagent gas than sample, most of the emitted electrons collide with reagent gas molecules, forming reagent ions. These reagent gas ions react with each other, in primary and secondary reaction processes that establish an equilibrium. They also react in various ways with sample molecules to form sample ions. CI ion formation involves much lower energy and is, therefore, much more "gentle" than electron ionization. Because CI results in much less fragmentation, CI spectra usually show high abundance of the molecular ion. For this reason, CI is often used to determine the molecular weights of sample compounds.


    Mass Filter
    In the mass filter, the ions are separated based on their mass-to-charge ratio (m/z). There are four classes of mass filters (analyzers) that are used to select and filter ions:
  • Time-of-flight (TOF)
  • Radio frequency (both quadrupole filter and ion trap)
  • Fourier Transform: Ion Cyclotron Resonance (ICR or FTMS)
  • Magnetic sector
    Play the video Download the video (1.9 Mbytes) Return to top of page
    The mass filter allows selected ions to move from the ion source to the detector. The mass analyzer filters ions based on their mass-to-charge ratio (m/z). Only the ions that match the mass filter parameters are passed through the mass filter at a given time.


    Detector
    The detector counts the ions and a signal is generated that is proportional to the total number of ions. Once an ionized sample has been filtered by mass-to-charge ratio in the mass filter, the ions are detected and reported to the data system.

    A detector is used to collect and count ions. Only one mass-to-charge ratio ion is detected at a given time whether the mass filter is a time of flight, separation by space, quadrupole filter or ion trap. Once inside the detector, the ions generate a signal that can be displayed by the data system.

    The plot of the mass-to-charge ratio (m/z) of these ions, as a function of abundance, is a mass spectrum.

    Play the video Download the video (0.3 Mbytes) Return to top of page

    Data System
    The data system has three main purposes:
  • Control the operational parameters of the ion source, mass analyzer and detector
  • Graphically display the results of the sample compound as a mass spectrum
  • Manipulate data for identification and quantification
    Play the video Download the video (0.5 Mbytes) Return to top of page
  •  
     