| Orthogonal-flow sprayer with pneumatic nebulizer |
Enhances sensitivity across a wide range of LC conditions and reduces maintenance requirements. |
| High-capacity solvent drying system |
Reduces chemical noise and permits the analysis of thermally-sensitive compounds through user control of gas flow rate and gas temperature. |
| Dielectric platinum-plated capillary |
Enhances sensitivity, permits independent optimization of the spray chamber and ion optics, and allows dynamic optimization of in-source fragmentation. |
| Air-cooled turbomolecular pump |
Provides fast startup and a small footprint, and eliminates the need for cooling water. Single, split-flow turbo pump enhances performance and increases reliability. |
| Fraction collection capability |
Can monitor and trigger the collection of fractions based on mass, an analog signal input, or time. For mass-based determination, up to 16 m/z values can be monitored. Requires optional I/O Accessory Module. |
| VL |
m/z 10 - 1500 |
| SL |
m/z 2 - 3000 |
| ± 0.13 u within the calibrated mass range in scan mode |
| VL |
2500 u/s |
| SL |
2500 u/s standard mode |
| |
5250 u/s fast scan mode |
ESI at 400 µ/min or APCI at 1000 µ/min |
| Selected ion monitoring of m/z 609.3 |
| Positive ionization |
| VL |
10 pg reserpine |
100:1 RMS (20:1 peak-to-peak) |
| SL |
1 pg reserpine |
100:1 RMS (20:1 peak-to-peak) | Note: When using the Multimode source in ESI mode, the same sensitivity applies. When using the Multimode source in APCI mode, the sensitivity is 50:1 RMS (10:1 peak-to-peak). In mixed mode operation a test mix at installation will provide no less than 20 % of the response obtained with dedicated sources.
ESI at 400 µ/min or APCI at 1000 µ/min |
| Scan range m/z 100 – 650 |
| Scan speed 2500 u/s |
| Extracted ion at m/z 609.3 |
| Positive ionization |
| SL |
50 pg reserpine |
100:1 RMS (20:1 peak-to-peak) |
| *There is no scan specification for the VL model |
| ESI |
Up to 1.0 ml/min water without a flow splitter |
| APCI |
Up to 1.5 ml/min water without a flow splitter |
| Ability to cycle through four acquisition modes on a scan-by-scan basis within a single run. Up to four analysis modes can be viewed in the software window and can be linked together to facilitate processing and analysis. Modes of operation: positive/negative polarity switching, variable fragmentor voltage selection, and concurrent SIM/scan. |
| Temperature range |
15° to 35°C (59° to 95°F) |
| Relative humidity |
less than 95% (non-condensing) |
| Height |
58 cm (23 in) |
| Width |
74 cm (29 in) |
| Depth |
66 cm (26 in) |
| Weight |
63 kg (139 lb) |
| *Approximate maximum cabinet dimensions with G1947A APCI source attached but without external mechanical pump. Height and width slightly less with G1948A ESI source. | Descriptions and specifications supersede all previous information and are subject to change without notice.
|
|