1200 Series Autosamplers - Specifications 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1

1200 Series Autosampler - Specifications at a Glance

Analytical injection ranges

Settable injection volume range Precision of peak areas Injection system Configuration
Standard autosamplers
0 - 100 µL 0.25 % RSD from 5 - 100 µL 1200 Series standard autosampler Standard
Up to 1500 µL Typically variable volume < 0.5 % (injection volume <1000 µL)
0.5 %( injection volume > 1000 µL)
G1329A with multi- draw kit

G1329A#021 or G13113-68711
0.1 - 900 µL Typically > 0.5 % (injection volume <1000 µL)
0.5 %( injection volume 5 - 900 µL)
G1329A with extended injection range option G1329A#020 or G1363A
Up to 1800 µL < 0.5 % up to 1800 µL Plus multi-draw kit
Up to 5000 µL < 0.5 % (injection volume <2000 µL)
1 %( injection volume > 2000 µL)
Plus additional exchange of seat capillary (0101-0301)
High performance autosamplers
0.1 - 100 µL (sampling from vials and well plates possible <0.25 % RSD from 5 - 100 µL,
< 1 % from 1 - 5 µL
1200 Series high performance autosampler and high performance autosampler SL

1200 Series well-plate autosampler SL
Standard
G1367B and G1367C
Up to 1500 µL (sampling from vials and well plates possible Typically variable volume, 0.5 % (injectioin volume, 1000 µL),
0.5 % (injection volume . 1000 µL)
G1367B/G1367C with multi-draw kit
G1367C#021 or G1313-68711

Extended injection ranges

Micro well-plate autosampler
0.01 - 8 µL,
0.01 - 40 µL
Typically < 0.5 % RSD from 5 – 40 µL,
< 1 % RSD from 1 – 5 µL,
< 3 % from 0.2 – 1 µL
1200 Series micro well-plate autosampler Standard (8-µL loop),
Extended loop capillary (40-µL loop), G1377A
Preparative autosamplers
0.1 - 900 µL
Up to 2300 µL
Typically < 0.5 % RSD from 10 – 500 µL,
< 3 % RSD from 1 – 10 µL,
< 0.5 % injection volume < 2000 µL),
1 % (injection volume > 2000 µL)
1200 Series preparative autosampler Standard
G2260

With multi-draw kit
Up to 5000 µL < 0.5 % injection volume < 2000 µL),
1 % (injection volume > 2000 µL)
1200 Series preparative autosampler With additional exchange of seat capillary (0101-0301)
Up to 50000 µL 1. Complet loop fill: 50 µL loop, overfill factor > 3: typically < 0.5 % RSD
2. Partial loop fill: 5000 µL loop, 1000 µL injection volume, draw/eject speed 20 mL/min: typically <1.0 % RSD
1200 Series dual-loop autosampler PS G2258A

 
 