Flow diagram for on-line 2D Nano LC/MS with 10-port valve 

The on-line 2D Nano LC/MS solution consists of a unique instrument configuration:

  • The high performance micro autosampler is connected directly to the quaternary pump.
  • From the pump the hydraulic flow passes through the micro 6-port/2-position valve of the micro well-plate autosampler and then the sample loop and needle.
  • The SCX column is positioned after the injector seat capillary and the SCX column outlet flow is directed to the 6-port valve of the autosampler.
  • This valve is connected with a capillary to the micro 10-port/2-position valve in the separate valve box.
  • Sample eluting from the SCX column is concentrated on a C18 enrichment column, which is mounted between two ports of this second valve and the eluting flow goes to waste.
  • The nanoflow pump is also connected to this second valve and the flow from this pump passes through the analytical column into the Agilent 6500 Series Accurate Mass Q-TOF LC/MS.
  • Peptides eluted at the current salt step concentration are trapped on the enrichment column, then the enrichment column is switched into the nanoflow path.
  • The nanoflow pump then delivers the elution gradient and the peptides are separated on the reversed-phase nano column.

Flow diagram for on-line 2D Nano LC/MS with 10-port valve

 

 

 

 

 

 

 

 

 

 

Back to On-line 2D Nano LC/MS

 
 