RNA 6000 Nano Chip Kit

Lab-on-a-Chip | RNA 6000 Nano Chip kitThe RNA 6000 Nano Chip kit provides information on the quality and quantity of RNA samples. The assay kit is designed primarily to check the quality and determine the concentration of total RNA and mRNA samples. For the analysis and quantitation of RNA preps, the RNA assay kit will be useful to scientists analyzing samples in the context of:
• Total and mRNA purification
• Northern blot analysis
• Analysis of RT-PCR products
• Expression profiling using DNA arrays
• In-vitro RNA transcription assays
• Preparation of cDNA libraries

The RNA 6000 Nano Chip kit has become the industry standard for the quality control and quantitation of total and messenger RNA samples. It contains a lower marker, which allows for sample alignment and permits easy comparison of samples enabling researchers to distinguish different types of mRNA based on the electrophoretic traces. Most importantly, the RNA 6000 Nano Chip kit provides enhanced RNA quantitation simultaneous with the assessment of RNA quality. Now, researchers can forego in many cases UV or ribogreen measurements and get both quality control as well as a good concentration measurement in one single analysis. Relative standard deviations of 10 % can be achieved, while the method is much less susceptible than UV to sample contaminants such as Phenol or genomic DNA. The RNA 6000 Nano Chip kit allows detection of total RNA down to 5 ng Digital data management and a series of software features further strengthen the leading position of the Agilent 2100 bioanalyzer for the analysis of RNA samples.


RNA 6000 Nano Chip Kit analytical specifications
Total RNA mRNA
Qualitative range 5-500 ng/µl 25-250 ng/µl
Quantitative range 25-500 ng/µl 25-250 ng/µl
Reproducibility of quantitation 10 % CV 10 % CV
Maximum sample buffer strength 10 mM Tris-EDTA 10 mM Tris-EDTAs
.
RNA 6000 Nano LabChip Kit physical specifications
Analysis run time 30 minutes
Number of samples 12 samples/chip
Sample volume 1 µl
Assay kit stability 3 months at 4 °C