采用安捷伦3000 性能增强型微型气相色谱仪检测天然气中的硫气味剂
Abstract:
采用安捷伦3000 性能增强型微型气相色谱仪,在3 min 内分析了天然气中低浓度(10 to 100 ppm) 的硫气味剂。Micro GC 配备四个并行通道:带反吹的PLOT U , OV-1 和两个OV-1701。在设定的条件下,硫的检测限约10 ppm。该方法也适于分析水源样品中的硫化氢。H2S 可检测至15 ppm 以下。
Analysis of natural gas and sulfur odorants at low levels (10 to 100 ppm) is realized in less than 3 minutes by the performance-enhanced Agilent 3000 Micro GC. The Micro GC was equipped with four parallel channels: PLOT U with backflush, OV-1, and two OV-1701s. The detection limits of sulfurs are around 10 ppm at the set conditions. This method is also suitable for the analysis of hydrogen sulfide wellhead samples; H2S can be detected down to 15 ppm.
Keywords:
sulfur odorants; natural gas; performance enhanced Micro GC, Heteroatom Compounds
硫气味剂,天然气,性能增强微型气相色谱,杂环化合物
Publication Number:
5989-9250CHCN
Last Updated:
11/2/2009
Number of Pages:
4