采用安捷伦双等离子体硫化学发光检测器和7890A 气相色谱仪依据ASTM D5623 检测汽油中的含硫化合物

Abstract: 依据ASTM D5623,采用Agilent 7890A 气相色谱仪配置355 双等离子体硫化学发光检测器,检测汽油样品中的含硫化合物。355 SCD 线性,等摩尔响应,可提供从0.1 至10 ppm 的线性范围,烃类和硫峰的共流出不干扰检测器的测定。汽油中含硫化合物的检测限可低至20 ppb,且在汽油样品分析中,无淬灭现象。 Sulfur components in gasoline samples are detected utilizing ASTM D 5623 with an Agilent 7890A GC configured with Agilent 355 dual plasma sulfur chemiluminescence detector (DP SCD). The 355 SCD response is linear, equimolar, and provides a linear range from 0.1 to 10 ppm. The coelution of hydrocarbon and sulfur peaks does not present a problem for the detector. The detection limits of sulfur compounds in gasoline are down to 20 ppb, while no quenching was found in the gasoline sample analysis.


Keywords: Dual plasma, sulfur, chemiluminescence, sulfur chemiluminescence detector, SCD, Sievers, Gasoline, Light hydrocarbons, fuel, sulfur compound, ASTM , 5623. 双等离子体,硫,化学发光,硫化学发光检测器,SCD,汽油,Sievers,汽油,轻烃,燃料,硫化合物,ASTM,5623