Two-Dimensional Gas Chromatographic Analysis of Trace Benzene in Styrene
Abstract:
A two-dimensional gas chromatographic method for the
analysis of trace benzene in styrene is described. This
method employs the Agilent 6890N GC system equipped
with a Deans switch device. The method can be used in a
prefractionator mode to analyze trace benzene in styrene
according to the proposed ASTM method [1]. It can also
be used in a heart-cutting mode to transfer more precisely
just the benzene peak from the primary column to the secondary
column. The Agilent 6890N electronic pneumatics
control and flow calculation software are used for setting
up the precise flows and pressures required. Excellent
calibration and sensitivity of this system is demonstrated
in both prefractionator mode and heart-cutting mode, and
the interference can be greatly reduced by the
heart-cutting.