Online literature is provided as Adobe Acrobat Portable Document Format (PDF) files. Obtain your free copy of the current Acrobat Reader from Adobe Systems.

Determination of Low Level Hydrocarbon Impurities in Propylene Using the Agilent 6820 Gas Chromatograph

Abstract: A method for analyzing trace hydrocarbon impurities in propylene is described. The method employs an Agilent 6820 gas chromatography (GC) system configured with a gas sampling valve, split/splitless inlet, and flame ionization detector. The Agilent Cerity Networked Data System for Chemical QA/QC was used to control the 6820 GC and to provide data acquisition and data analysis. An Agilent HP-Al2O3 column was used for separation of the trace hydrocarbons. Impurity levels at 1 ppm were easily detected in propylene. This method does not determine all possible impurities such as CO, CO2, H2O, alcohols, nitrogen oxides, and carbonyl sulfide, or hydrocarbons larger than decane.


Keywords: Propylene, 6820, petrochemical, PLOT, HP-Alumina, hydrocarbon impurities, Petrochemicals
Publication Number: 5988-9258EN
Last Updated: 11/2/2009
Number of Pages: 6

For your convenience this PDF file will open in a separate window, and you may need to adjust your pop up blocker setting.