Agilent 6850 Sub-Ambient Oven Performance for ASTM D3710 Simulated Distillation of Gasoline
Abstract:
The Agilent 6850 gas chromatograph is now offered with sub-ambient column oven capabilities using liquid carbon dioxide as the cryogenic gas. This extends the 6850 oven temperature from -20C to 350C. The sub-ambient oven performance of the 6850 was tested using the ASTM D3710 method for simulated distillation of gasoline range materials. This method requires starting oven temperatures of -20C followed by programming to 180C. Four chromatographic performance tests required by the ASTM method were used to evaluate 6850 system performance. The same tests were also run on the Agilent 6890 GC to offer a comparison with an industry standard system. Both the 6850 and 6890 results exceeded the specifications for peak shape, resolution, and retention time repeatability and area percent repeatability. In most cases, both GCs showed performance that was 10 times better than the ASTM requirements. The 6850 and the 6890 also showed nearly identical performance when compared to each other. Additionally, the 6850 used about one-half of the CO2 cryogenic gas compared to the 6890 when running this method.
Keywords:
simdis, D3710, cryo, cryogenic, 6850, ASTM, gasoline, Simdis
Publication Number:
5980-0250E
Last Updated:
11/2/2009
Number of Pages:
6