System |
|
|
Flow range |
0.1 mL/min to 5 mL/min (settable), 1 mL to 5 mL/min recommended |
|
Maximum operating pressure |
600 bar |
|
Upgrade possibility of existing 1100/ 1200 / 1260 LC |
yes |
|
Option for SFC/ UHPLC in one system |
yes |
|
CO2 purity requirements |
99.99% for gas, 99.999% for liquid |
|
Unattended operation |
Leak sensors, diagnostic software features |
|
Fusion A5 SFC Module |
|
|
backpressure accuracy |
+/- 1% |
|
Backpressure precision |
+/- 0.5 bar (+/-0.2 bar typical) |
|
Backpressure thermal precision |
+/- 1C |
|
Chiller temperature |
-20C to 9 C |
|
Pump |
|
|
Number of co-solvents |
1, with SSV pump option 3, with external SSV 12 |
|
Precision of flow rate |
Same as G1312B, ≤0.07 % RSD or ≤0.02 min SD, whatever is greater based on retention time at
constant room temperature |
|
Precision of composition |
Same As G1312B, < 0.15 % RSD or < 0.04 min SD whatever
is greater |
|
Autosampler |
|
|
Number of samples |
100 x 2 mL vials, 30 x 6 mL vials |
|
Injection volume |
5 uL for full loop injections (5 uL loop as default) larger loops possible (10uL), partial loop filling for smaller injection volumes |
|
Injection principle |
Fixed loop |
|
Injection precision |
<0.3% RSD for 5 ul |
|
Sample temperature control |
4-40C with sample thermostat |
|
Sample carryover |
0.05% |
|
Column compartment (TCC) |
|
|
Column |
Up to 300 mm length x 4.6 mm ID (or less) |
|
Column capacity |
Up to 3, with additional 2x TCCs up to 9 |
|
Switching valve |
Optional (2/6, 2/10, 8/9, 6 column selector) |
|
Temperature range |
Ambient -10C to 100 C |
|
Column compartement temperature accuracy |
+/- 0.5C |
|
Column compartement temperature stability |
+/- 0.05C |
|
Active solvent pre-heating and post-conditioning |
As standard |
|
Automated Method development |
optional |
|
Detector |
|
|
Wave length range |
190-950 nm |
|
Wavelenghth accuracy |
+/- 1 nm |
|
Linearity range |
2.0 AU @ 265 nm |
|
Noise, wet |
+/- 0.05 mAU under SFC conditions at 254 nm 20% MetOH/80% CO2
+/- 0.007 mAU under LC condition |
|
Drift |
0.9 mAU/hr |
|
Data rate |
80 Hz |
|
Cell pressure limit |
400 bar |
|
Cell path length |
10 mm |
|
Cell volume |
13 ul |