These videos can help you perform common maintenance procedures for an Agilent 5975 Series GC/MSD. As always, Agilent's Technical Support Team is available to answer any questions you may have regarding installation or maintenance of Agilent products.
Before you start any of the tasks below, please review the important safety precautions on pages 17 and 18 in the 5975 MSD Operation Manual found on the Instrument Utilities CD that was delivered with your system.
Installing the GC Column
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Open (flip down) the analyzer cover
- Open the analyzer
- Install the column into the transfer line
- Close the analyzer
- Turn on the MSD
- Pump down the MSD
- Close (flip up) the analyzer cover
- Install the front window cover
Replacing the Filaments
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Open (flip down) the analyzer cover
- Remove the analyzer
- Remove the source from the analyzer
- Remove and replace the filaments
- Install the source in the analyzer
- Install the analyzer
- Turn on the MSD
- Pump down the MSD
- Close (flip up) the analyzer cover
- Install the front window cover
Cleaning the EI Ion Source
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Open (flip down) the analyzer cover
- Remove the analyzer
- Remove the source from the analyzer
- Remove the filaments
- Identify ion source parts
- Source parts that need to be cleaned (also available as a downloadable and printable document, 541 kb)
- Disassemble the EI source
- Clean the EI source
- Sonicate the source parts
- Reassemble the EI source
- Install the filaments
- Install the source in the analyzer
- Install the analyzer
- Turn on the MSD
- Pump down the MSD
- Close (flip up) the analyzer cover
- Install the front window cover
Cleaning the CI Ion Source
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Open (flip down) the analyzer cover
- Remove the analyzer
- Remove the source from the analyzer
- Remove the filaments
- Source parts that need to be cleaned (pdf, 541 kb)
- Disassemble the CI source
- Clean the CI source
- Sonicate the source parts
- Reassemble the CI source
- Install the filaments
- Install the source in the analyzer
- Install the analyzer
- Turn on the MSD
- Pump down the MSD
- Close (flip up) the analyzer cover
- Install the front window cover
Replacing the Detector Electron Multiplier Horn
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Open (flip down) the analyzer cover
- Remove the analyzer
Note: If you have a triple-axis detector, skip to step 6.
- Remove and replace the horn from the analyzer
Note: The following steps should be done if you have a triple-axis detector. Otherwise, skip to step 8.
- Remove the triple-axis detector horn
- Install the triple-axis detector horn
- Install the analyzer
- Turn on the MSD
- Pump down the MSD
- Close (flip up) the analyzer cover
- Install the front window cover
Filling the Calibration Vial
- Remove the front window cover
- Remove, refill, and install the calibration vial
- Install the front window cover
Changing the Mechanical (Roughing) Pump Oil
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Drain the oil from the pump
- Add oil to the pump
- Turn on the MSD
- Pump down the MSD
- Install the front window cover
Changing the Diffusion Pump Oil
- In the MSD ChemStation or from the GC keyboard, initiate a vent and wait for the process to complete
- Turn off the MSD
- Remove the front window cover
- Open the vent valve
- Remove the GC column from the transfer line
- Move the MSD away from the GC
- Remove the diffusion pump
- Drain the oil from the pump and refill
- Install the diffusion pump
- Move the MSD back up against the GC
- Install the column into the transfer line
- Turn on the MSD
- Pump down the MSD
- Install the front window cover