KB000810: Can the Multiple Affinity Removal Column Capacity be Determined? - Multiple Affinity Removal System 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • Can the Multiple Affinity Removal Column Capacity be Determined?
      Yes, we recommend using human serum albumin (HSA) at 20mg/mL concentration in Buffer A (part number 5185-5989 is a ready-to-use HSA standard) for checking column capacity only when the peak heights (flow-through and eluted) become abnormal. Inject 30-45µL of this standard for the 4.6 x 50 mm column; 60-90µL for the 4.6 x 100 mm column.