Select a Country or Area
|
Contact Us
Products & Services
Technical Support
Solutions
Buy
About Agilent
Go to Part Number Search
Home
>
Technical Support
>
FAQs
>
Mass Spectrometry
>
Software
>
MS ChemStations
Register
|
Login
KB005168: What does Error Message "MS in use by another user" Indicate? - MS ChemStation
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying
What does Error Message "MS in use by another user" Indicate?
The Server Error 32 and MS in use by another user errors occur with G1701DA Rev. D.01.02 and G1701DA Rev. D.02.00
This is directly related to the number of times the MSD chemstation is started in relation to the power cycles of the MSD.
With G1701DA Rev. D.01.02 and G1701DA Rev. D.00.02, it is observed that after the 36th startup of chemstation, the Server 32 error is reported and after 19th startup of the chemstation the MS is in use by another user error will be reported.
To avoid these errors, power cycle the MS before the critical number of chemstation restarts. This should be done once or twice a week.