Yes. The inactivity timeout can be configured with G1732AA rev A.02.00. To set the Time-Based Session Lock, follow the steps below:
- Right-click anywhere on the desktop and select Properties from the context menu. In the Display Properties dialog box, select the Screen Saver tab. The time-based session lock is managed by the special Agilent GCMS21CFR11 screen saver.
- Change the Wait time to 1 minute and click OK to close the Display Properties dialog box.
- Wait for 1 minute, the Session Lock dialog box is displayed, and the workstation is locked.
- Enter the appropriate UserID and Password to unlock the workstation, then open the Display Properties dialog box again and reset the Wait time to its previous value.
- Click OK to close the Display Properties dialog box.