  • Why are No Peaks Identified After Replacing the Column for SIM Analysis?
      SIM analysis typically performs grouping and changes the ions captured in each time band. Changing the column usually causes the peak retention times to change. If the retention time for a particular compound's peak changes the peak will not appear if the ions required to capture that compound's peak are not set. In this situation, perform another scan measurement, reconfigure the SIM data acquisition conditions based on this data, and then perform the SIM measurement. You can also obtain the data by performing synchronous SIM/Scan data acquisition.