KB001398: What are the Steps to use Manual Integration to Update a Calibration Table? - MS ChemStation 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • What are the Steps to use Manual Integration to Update a Calibration Table?
      The G1701XX software saves the quantitative results in to a Quantresult file. The file is located in the data file directory, has the name of the method and the extention is .RES.
      To get a quant-report with a manually integrated peak, perform the following steps:
      1. Create a quant report
      2. Select in Data Analysis [View] + [QEdit Quant Results]
      3. Select the peak from the "Quick QEdit" compound list
      4. From the menu select [Display] + [Qualifier]
      5. Do the manual integration
      6. Close "Quedit Quant Results" and save the result.
      7. Back in the Data Analysis View select [Quantitate] + [Generate Report]. The response of manual integreted peaks will be marked by a "m".
      8. Select from the menu [Calibrate] + [Update...] + update one level.
      9. Answer the following question: File has previuosly been quantitated. Requantitate now? with NO!
      10. Select the update mode from the next screen and press the button [Do Update]
      11. Check on page 3 of the calibration table to verify the response is correct.