Agilent Technologies LSCA - Our measure is your success
clear
white
clear
white
clear

KB001299: What does Error Message "Server 32 error" Indicate? - MS ChemStation 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • What does Error Message "Server 32 error" Indicate?
      The Server Error 32 and MS in use by another user errors occur with G1701DA Rev. D.01.02 and G1701DA Rev. D.02.00
      This is directly related to the number of times the MSD chemstation is started in relation to the power cycles of the MSD.
      With G1701DA Rev. D.01.02 and G1701DA Rev. D.00.02, it is observed that after the 36th startup of chemstation, the Server 32 error is reported and after 19th startup of the chemstation the MS is in use by another user error will be reported.
      To avoid these errors, power cycle the MS before the critical number of chemstation restarts. This should be done once or twice a week.
grey1
related_bottom_left clear clear clear
clear clear
grey2
Please select your country:
 
   
Registration   |  Login Benefits of Registration