  • What does Error Message "MASS SPEC IN USE BY ANOTHER SESSION" Indicate?
      Typically you can recover from this error by following the steps below: Exit the ChemStation and reboot the PC. Do not start the ChemStation at this time. Next, the Smartcard in the 5973N MSD will need to be reset. Attempt this by using the display on the MSD (LUI) by pressing Menu repeatedly until Conroller menu is displayed. Next, press the Item button until Reboot Controller? is displayed. Select the Yes/Select key. If this fails to reboot the MSD controller, then the MSD must be powered off and then back on. Allow three to five seconds (no more) before powering the MSD back on. Secondly, if you have added a second lan card, follow the instructions for setting the bindings for the lan card on one of the document links below: