KB000454: What does Error Message "Invalid path C:\MSDCHEM\MSEXE\DASTATE.MAC" Indicate? - MS ChemStation 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • What does Error Message "Invalid path C:\MSDCHEM\MSEXE\DASTATE.MAC" Indicate?
      The DASTATE.MAC file can become corrupt when the disk is full. This file is easily repaired as follows:
      1. Locate C:\MSDCHEM\MSEXE\DASTATE.MAC (or C:\HPCHEM\MSEXE\DASTATE.MAC) and open in Notepad. (if this file does not exist, open Notepad and skip to step 3)
      2. Clear all contents of this file
      3. Create the following 2 lines only:
           NAME LOADSTATE
           REMOVE LOADSTATE
      1. Save this file as C:\MSDCHEM\MSEXE\DASTATE.MAC (or
      2. C:\HPCHEM\MSEXE\DASTATE.MAC for versions G1701BA and earlier).
      3. Start Data Analysis and then Exit
      4. Exiting data analysis should write the additional lines this file normally holds.
      DASTATE.MAC remembers the state of various options when you close Data Analysis, so they are set correctly when you open Data Analysis later.