Why do the Ion Ratios Not Change when Using QEDIT?
The
Subtraction Method
must be set to
Extended Area Quant
. In both Enhanced and EnviroQuant mode of G1701xx, it is possible to integrate each of the extracted ion chromatograms used for quant and qualifier signals independently of one another. To do this, the compound must be set for extended area quant by selecting
Extended Area Quant
as the
Subtraction Method
for the compound(s) on page one under
Edit Compounds
. To edit individual ion chromatograms, select
Qedit
, then
Display
and then select
Qualifiers
. Ion chromatograms are then displayed in a separated format, rather than merged. It is possible to manually integrate any and all of the ion chromatograms displayed. Updated quant ion areas are displayed in the report, along with corrected ion ratios.