Agilent Technologies LSCA - Our measure is your success
clear
white
clear
white
clear

KB000606: Is Color Compensation Necessary? - LOAC Cell Assays 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • Is Color Compensation Necessary?
      No, fluorescence color compensation is not needed. The two detection channels are far from each other (525 and 680 nm), so that overlap of emission spectra is very unlikely.
grey1
related_bottom_left clear clear clear
clear clear
grey2
Please select your country:
 
   
Registration   |  Login Benefits of Registration