KB000183: What does Error "Quad DC Fault" During the MSD Ion Optics Test Indicate? - LC/MS ChemStation 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • What does Error "Quad DC Fault" During the MSD Ion Optics Test Indicate?
      With the LC/MSD ChemStation rev. A.05.02/.04 software, the MSD Ion Optics Test in Diagnosis will often give a "Difficulty with ion optics electonics... Quad DC Fault" when it does the 'On Maximum (Neg.)' part of the test. This is a problem with the test, so don't troubleshoot an instrument problem because of this particular failure of the test. This fault in the Ion Optics test will be corrected in rev. A.06.01.