KB004906: What are the Considerations When Using the ELSD with Other Detectors? - 1200 LC 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • What are the Considerations When Using the ELSD with Other Detectors?
      As a consumptive detector, the ELS detector is the last one in the flow path or can be used with a splitter in combination with other consumptive detectors. The standard nebulizer used at a flow rate of 1 ml/min generates a back-pressure of 4 bar, which is compatible with most other flow cells. Depending on the selected nebulizer, a back pressure of up to 44 bar is generated for high flow rates, which is not compatible with some flow cells, see respective user manuals for details else contact Agilent Technologies for details.