KB005112: How do I configure a column from the ChemStation? - 7890A GC 

white
clear clear clear clear clear

grey1
Downloads and Utilities
Education & Events
Frequently Asked Questions
DNA MicroArrays
Gas Chromatography
Lab-on-a-Chip
Life Sciences Informatics
Liquid Chromatography
Mass Spectrometry
Networked Data Systems
Proteomics
Installation and Maintenance Videos
Literature Library
Parts Information
Description, Pricing, and Availability
Illustrated Parts Breakdowns
Support Services - Finding
Warranty Status - Verifying

grey1

  • How do I configure a column from the 7890A GC ChemStation?

      This document explains how to configure a column from your Agilent software for 3 versions of the ChemStation software: B.03.01, B.03.02, and B.04.01. The screens found in the MSD ChemStation and EZChrom software are almost the same.Once you have launched the software, select GC Configuration menuItem from the Instrument menu.

        The Configuration screen that appears has two sets of tabs.

      1.      Select the Configration tab.

      2.      Select the Columns tab.

      Column Configuration – version B.03.01

      To Install a column, click the button next to a column row to add (+) or remove (-) a column from the table.  Clicking the button under the +/- column heading brings up the screen below to install the selected column, calibrate selected column, or delete a column.

      This feature was not implemented completely – there were three major problems:

      ·         The column numbers were not displayed on the interface

      ·         Once a column was added, it could not be removed.

      ·         The lowest Temperature Limit for any column, even if it was not being used, was applied to the method.

      Column Configuration – version B.03.02 and B.04.01

      In the recent version of ChemStation, the columns are numbered, and can be added and removed. For step by step instructions, see below.

      1.  Select the Columns tab to configure columns, calibrate the column, and select the inlet and output points. Up to 6 columns can be configured and each column is in order from 1 to 6.

      2.  Highlight by selecting the column description area you want to install, such as column 1 as shown above.  Select the Inventory button.

      3.  Select the Add Column to Local Inventory button.

      4.  Scroll down and select the J-413 Checkout column.  Enter a unique inventory # or use the number provided.  The New Inventory# is generated based on date and time so it is unique; Users can enter the Serial # from the column tag for example.

      5.  Select the Add Selected Column to Inventory button.

      6.  Select Install Selected Column.

      7.    Select the Inlet, Outlet, and Heated By drop-down arrows and make the selection.

      8.      From this screen, the column can be calibrated or removed.

      9.      The up/down arrows are used to move for example, column in position 2 to the 4th column position.



* required field

Did this article:
Provide the information you were looking for?  


Help us help you:
Suggestions to improve this document.
Note: If you would like a return response, please include your email address




grey1
related_bottom_left clear clear clear
clear clear
grey2
Please select your country:
 
   
Registration   |  Login Benefits of Registration