Agilent Technologies has tested certain ODBC driver libraries for compatibility with ChemStore. The use of any other libraries risks the possibility of corrupting data that is in the database. Therefore, Agilent only recommends using the specifically supported MDAC (Microsoft Data Access Components) libraries suited for your specific version of ChemStore and Security Pack.
The most common cause of this error is that either new software (e.g. MS Office or an operating system service pack) has been installed on your system which has replaced the supported MDAC version of the library.
For systems with revision B.03.03 or higher of ChemStore, this error is triggered as a warning only. The warning can be acknowledged and the application starts. If you like to suppress the warning a check box allows you to suppress it from that time on.
For systems with revision B.02.01 to B.03.02 of ChemStore, you must modify the correct registry keys
(HKEY_LOCAL_MACHINE > SOFTWARE > Hewlett-Packard > HP ChemStore CS > OdbcDriverAccess and OdbcDriverOracle) to the new revision numbers. This allows the software to start up successfully but using an unsupported revision of MDAC.
The following table describes which version of ODBC driver libraries are supplied with which version of operating system.
Note: These are the only supported versions of the drivers for use with ChemStation!
| Component
| ACCESS ODBC Driver Ver.
| ORACLE ODBC Driver Ver. |
| Windows 2003 R2 SP2
| 04.00.6305
| 10.02.00.03 or 10.02.00.04 |
| Windows 2003
| 04.00.6205
| 02.575.1022 |
| WinXP SP3
| 04.00.6305
| 10.02.00.03 or 10.02.00.04 |
| WinXP SP2*
| 04.00.6304
| 02.575.1117 |
| WinXP SP1
| 04.00.6019
| 02.575.1022 |
| Win2K SP3,4
| 04.00.6200
| 02.573.6200 |
| MDAC 2.8 Patch 927779
| 04.00.6200
| 02.575.1064 |
| MDAC 2.8
| 04.00.6200
| 02.575.1022 |
| MDAC 2.7 RTM |
04.00.6019 |
02.573.9001 |
| MDAC 2.6 |
04.00.6019 |
02.573.6526 |
| MDAC 2.5
| 04.00.6019
| 02.573.6019 |