Software Status Bulletin
G6854AA MassHunter Personal Pesticide Database rev. B.02.01
Revision Log:
|
Revision Date |
Reason For Update |
|
11/25/2008 |
Final release |
|
4/16/2009 |
Updated for rev. B.02.01 |
1. Overview
The purpose of this document is to list known problems in the software and any known workarounds or recovery paths.
2. Known Problem Summary
The following is a summary of the known problems in the software. This is not an exhaustive list, but is a list of the problems a customer is most likely to encounter. This list cross-references items in our Defect tracking database.
|
SCR # |
Title |
|
11989 |
Pesticides.mtl is only accessible by the user that installed PCD. |
|
12437 |
Extra page in report when "print structures" is selected. |
|
13248 |
.mtl file should be called by generic name in Qual |
|
14081 |
Structures can sometimes not be legible in printouts. |
3. Known Problems Detailed Description
The following is the detailed descriptions of the known problems in the Personal Pesticide Database software. This section includes more extensive information from our defect tracking database. This section also includes any known workarounds or recovery paths.
|
SCR #11989 |
|
Title |
Pesticides.mtl is only accessible by the user that installed PCD. |
|
Description |
To reproduce this problem:
Install PCD as User1. Login in as User2. The Pesticides.mtl file is read-only for that User2 so it doesn't work.
Workaround:
Use the Windows File Explorer to change the permissions for all the PCD (.mtl) database files to be write-able (not read-only) by all intended users.
These files are installed under the MassHunter "home" directory, under "databases". For example: D:\MassHunter\databases |
|
SCR #12437 |
|
Title |
Extra page in report when "print structures" is selected. |
|
Description |
To reproduce:
-run a single search -highlight 3 compounds in search results -right-click on results section and select Print Results -select "landscape", "fit to page", "only highlighted items", "include search parameters" and "print structures" -select 360X240 or 540x360 and print.
The report has 3 pages; page 3 is empty except for the column headers. |
|
SCR #13248 |
|
Title |
.mtl file should be called by generic name in Qual |
|
Description |
In Qualitative Analysis, the Open Database File dialog allows the user to choose files of type .csv or .mtl. The .mtl file type is called "Metlin database file". This should have a generic name in case the customer is using a different personal database product. |
|
SCR #14081 |
|
Title |
Structures can sometimes not be legible in printouts. |
|
Description |
When printing structures and using Fit to Page the structures can end up being too small to see.
Workaround:
Print in Landscape mode. |
Appendix A. Glossary
|
Term / Acronym |
Definition |
|
UI |