Software Status Bulletin
G3785AA LC/SQ ChemStation Integration Software rev. B.01.00
Revision Log:
|
Revision Date |
Reason For Update |
|
4/2/2008 |
Final release |
1. Overview
The purpose of this document is to list known problems in the software and any known workarounds or recovery paths.
2. Known Problem Summary
The following is a summary of the known problems in the software. This is not an exhaustive list, but is a list of the problems a customer is most likely to encounter. This list cross-references items in our Defect tracking database.
|
SCR # |
Title |
|
12681 |
If destination is blanked and then translate, translation happens to last destination. |
|
12710 |
Error when loading some translated MWD datafiles in Mass Hunter. |
|
12735 |
Profile data from ChemStation gets translated as peak detected data. |
3. Known Problems Detailed Description
The following is the detailed descriptions of the known problems in the LC/SQ ChemStation Integration software. This section includes more extensive information from our defect tracking database. This section also includes any known workarounds or recovery paths
|
SCR #12681 |
|
Title |
If destination is blanked and then translate, translation happens to last destination. |
|
Description |
To reproduce this problem:
1. Select a file to translate.
2. Enter a valid destination path
3. Select a valid path,e.g. Desktop
4. Highlight the path and delete it so that the field is empty
5. Without clicking anywhere else, click the Play or Translate button
6. Translation occurs to the Desktop and then an error is shown in the destination field
The translation should not occur and the error should note that the destination is incorrect. |
|
SCR #12710 |
|
Title |
Error when loading some translated MWD datafiles in MassHunter. |
|
Description |
For some MWD data created on the ChemStation, the first data record of the next file will be appended to the last data record of the current file. As a result, the current file will give an error upon being loaded into MassHunter. This problem is not easily reproducible. |
|
SCR #12735 |
|
Title |
Profile data from ChemStation gets translated as peak detected data. |
|
Description |
Profile data from ChemStation will get translated as peak detected data. |
Appendix A. Glossary
|
Term / Acronym |
Definition |
|
UI |