Software Status Bulletin
6400 Series LCMS Triple Quadrupole Bundle
MassHunter WorkStation Acquisition Software rev. B.02.01
Revision Log:
|
Revision Date |
Reason For Update |
|
27 May 2009 |
Finalized for B.02.01 release |
|
08 Jan 2009 |
Added SCR# 13496 as critical – cannot use MMI with 6460 BETA SW and MH Optimizer |
|
06 Jan 2009 |
Documents for B.02.01 BETA release |
|
09 Sept 2008 |
Document for B.02.00 release |
1. Overview
The purpose of this document is to list known problems in the software and any known workarounds or recovery paths.
2. Known Problem Summary
The following is a summary of the known problems in the software. This is not an exhaustive list, but is a list of the problems a customer is most likely to encounter. This list cross-references items in our Defect tracking database.
2.1 Installation
|
SCR # |
Title |
|
06752 |
Setting ALS SL thermostat to off does not work. |
|
9870 |
Uninstalling QQQ SW wipes out Instrument Actuals settings. |
|
12531 |
G1367B hALS Tray Configuration is initially unavailable. |
|
12698 |
Choose MassHunter Workstation Acquisition default installation location for program files. |
|
13549 |
MassHunter Workstation Acquisition cannot restart LC detector module if failure occurs. |
|
13551 |
MassHunter Workstation Acquisition does not start if configured detector becomes disconnected. |
|
13688 |
Unable to Change Parameters in Manual Tune In New Installations. |
2.2 Tuning
|
SCR # |
Title |
|
12677 |
Collision Cell Gas will not be turned on before Autotune and Checktune start. |
|
13169 |
Axis and width offset parameters settings sometimes stay dynamically ramped when Manual Tuning. |
|
13184 |
Manual ramping may use different "from" and "to" values than specified when Manual Tuning. |
|
13193 |
Default ramp value for prefilter DC2 is out of range when Manual Tuning. |
|
13262 |
HPLC-Chip autotune does not complete successfully. |
|
13519 |
Selecting scan stepsize 5 causes Lens2 DC driver fault in Manual Tune. |
|
13831 |
View Tune Parameters in MS QQQ tab in Acquisition context only displays positive Tune Parameters. |
2.3 Acquisition
|
SCR # |
Title |
|
8751 |
A worklist does not abort when LC slave module loses power. |
|
9450 |
MassHunter Acquisition blacks out when adjusting Real-time Plot pane. |
|
12696 |
Only 4 Scan Segments allowed when in MS2Scan Mode |
|
12517 |
Acquisition Method sporadically fails to Import into Quantitative Analysis. |
|
13230 |
Ion source field is blank at first startup for 6460 using ESI with Agilent Jet Stream Technology. |
|
13225 |
Error may occur when starting offline Acq Method and Worklist Editor when acquisition is online. |
|
13498 |
Solvent Manager fails if new solvent names have blank spaces in them. |
|
|
|
|
|
|
|
13496 |
Multimode Ion Source does not work with MassHunter Optimizer B.02.00. |
|
13498 |
Solvent Manager fails if new solvent names have spaces in them |
|
13687 |
Cannot Save Instrument Curves Settings at Next Startup. |
|
|
|
|
13689 |
Ion Source and parameters are blank when offline Method for ESI with Agilent Jet Stream Technology is loaded for online acquisition. |
|
13694 |
Ghost Peak Sometimes Appear at 3000 m/z. |
|
|
|
|
13933 |
Only Contact 1 is visible in External Contacts tab for G1367D ALS SL+. |
|
13979 |
Needlewash location "flushseat" changes to "as method" when method is saved |
|
13981 |
MassHunter does not handle blank spaces. |
|
16146 |
Injector Program AND Overlapped Injection in Same Method Does Not Work. |
2.4 Reports
|
SCR # |
Title |
|
13513 |
Wait for Autotune report to show before pressing OK when Autotune completes. |
|
13832 |
MassHunter Workstation Acquisition Method report generation sometimes hang when printing. |
3. Known Problems Detailed Description
The following is the detailed descriptions of the known problems in the QQQ software. This section includes more extensive information from our defect tracking database. This section also includes any known workarounds or recovery paths.
3.1 Installation
|
SCR #06752 |
|
Title |
Setting ALS SL thermostat to off does not work. |
|
Description |
Setting ALS SL thermostat to off prompts MassHunter to become not ready. The not ready message is “cooling samples”. |
|
Workaround |
Specify a temperature for the ALS-SL. . |
|
SCR #9870 |
|
Title |
Uninstalling QQQ SW wipes out Instrument Actuals settings. |
|
Description |
If a user uninstalls and reinstalls the QQQ software, the settings for the Instrument Actuals panel |
|
Workaround |
Right-click on the Actuals panel in the newly installed QQQ software to get to the Setup dialog. This will enable the user to put in new Actual settings. |
|
SCR #12531 |
|
Title |
G1367B hALS Tray Configuration is initially unavailable. |
|
Description |
On a G1367B hALS with A.06.10 firmware, MassHunter Workstation Acquisition cannot select the Tray Configuration. |
|
Workaround |
Do not connect APG remote cable on G1367B hALS: move to pump or detector LC modules. |
|
SCR #12698 |
|
Title |
Choose MassHunter Workstation Acquisition default installation location for program files. |
|
Description |
Starting with MassHunter Workstation B.01.04, any directory path for the acquisition software can be specified as the destination location for the program. Installation creates any directories that did not pre-exist and proceeds to completion. However, not all functionality is available. |
|
Workaround |
Choose MassHunter Workstation Acquisition default installation location for program files. |
|
SCR #13230 |
|
Title |
Ion source field is blank at first startup for 6460 using ESI with Agilent Jet Stream Technology. |
|
Description |
When the G6460A Triple Quad LC/MS first starts up using ESI with Agilent Jet Stream Technology, the ion source field is blank. Due to the new technology, ESI has to be chosen first as the ionization technique. |
|
Workaround |
Choose ESI for the ion source field, and ESI with Agilent Jet Stream Technology will be available. |
|
SCR #13496 |
|
Title |
Multimode Ion Source does not work with B.02.01 |
|
Description |
The "IonSource" field in the method has been modified in Acquisition to be "MultiMode" from "MMI". This makes the Optimizer incompatible with Multimode methods on B.02.01. |
|
SCR #13549 |
|
Title |
MassHunter Workstation Acquisition cannot restart LC detector module if failure occurs. |
|
Description |
If a LC detector (like VWD or DAD) loses power or get its LAN disconnected during a worklist acquisition, the module goes gray (offline), and the worklist stops. MassHunter Workstation Acquisition cannot turn on the LC detector again to a ready status. |
|
Workaround |
Restart MassHunter Workstation Acquisition is required if LC device power failure occurs or LAN gets disconnected (on the CAN bus). |
|
SCR #13551 |
|
Title |
MassHunter Workstation Acquisition does not start if configured detector becomes disconnected. |
|
Description |
If a configured LC detector (like VWD or DAD) get its LAN disconnected during startup, MassHunter Workstation Acquisition does not start. |
|
Workaround |
Restart MassHunter Workstation Acquisition is required once LAN gets re-connected (CAN bus). |
|
SCR #13688 |
|
Title |
Unable to Change Parameters in Manual Tune In New Installations. |
|
Description |
With B.01.04 and B.02.01, a new software installation does not allow changes in manual tune without starting an autotune first. If there is a no ion problem with the instrument, the first step in troubleshooting includes setting width gain to 100 and width offset to 4 ("opening" the quads). These do not work for new installations with no corresponding changes noted in the real time plot. |
|
Workaround |
Start an initial autotune first. Whether it passes or fails, changing parameters in Manual Tune now work. |
3.2 Tuning
|
SCR #12677 |
|
Title |
Collision Cell Gas will not be turned on before Autotune and Checktune start. |
|
Description |
Autotune and Checktune should properly be run with the Collision Cell Gas turned on. If the user turns the Collision Cell Gas off, and then initiates an Autotune or a Checktune, these programs will not turn it back on. If the user puts the instrument into Standby mode, and then turns it On, the Collision Cell Gas will be turned On. |
|
Workaround |
Do not turn the Collision Cell Gas off. There is no need to turn the Collision Cell Gas off. It uses very little gas (~2 ml/min at atmospheric pressure). When in Standby mode, the system automatically sets the Collision Cell Gas to one-half its factory set value. When the instrument is turned to On mode, the Collision Cell Gas is turned on. |
|
SCR #13169 |
|
Title |
Axis and width offset parameters settings sometimes stay dynamically ramped when Manual Tuning. |
|
Description |
In Manual Tune, disabling the dynamic ramp parameters for axis and width offsets does not stay. That is, axis and width offsets stay dynamically ramped even if tune file is saved as another file after disabling. |
|
SCR #13184 |
|
Title |
Manual ramping may use different "from" and "to" values than specified when Manual Tuning. |
|
Description |
In Manual Tune and as an example, select fragmentor to manually ramp from the parameter menu. Choose from 0 to 150 with stepsize 50. The fragmentor ramp starts at 30 and ends at 180. |
|
SCR #13193 |
|
Title |
Default ramp value for prefilter DC2 is out of range when Manual Tuning. |
|
Description |
The default "from" ramp value for prefilter DC2 is -60 V, and this value is out of range. It should be -50 V. |
|
Workaround |
Change prefilter DC2 “from” ramp value to -50 V. |
|
SCR #13262 |
|
Title |
HPLC-Chip autotune does not complete successfully. |
|
Description |
Autotune with HPLC-Chip positive does note complete successfully. |
|
Workaround |
Tune with ESI, ESI with Agilent Jet Stream Technology, or MMI prior to installing and using HPLC-Chip. |
|
SCR #13519 |
|
Title |
Selecting scan stepsize greater than 3 causes Lens2 DC driver fault in Manual Tune. |
|
Description |
In Manual Tune, selecting a stepsize larger than 3 produces no observed chromatogram peaks in the RTP for scan. |
|
Workaround |
Do not select stepsize greater than 3 for scan. |
|
SCR #13831 |
|
Title |
View Tune Parameters in MS QQQ tab in Acquisition context only displays positive Tune Parameters. |
|
Description |
In the Acquisition context and on the MS QQQ tab, selecting View Tune Parameters always show positive tune parameters only. This is regardless if the ionization mode is positive or negative. |
|
Workaround |
Check the Tune Parameters in the Tune context. |
3.3 Acquisition
|
SCR #8751 |
|
Title |
A worklist does not abort when LC slave module loses power. |
|
Description |
If an LC slave module (i.e. not containing the network card) is powered off during a worklist run, it only partially aborts. Abort requests appear in the system log, but are not executed. If the LC module is powered on again its device icon shows it as available, but you can neither stop nor abort the worklist. |
|
Workaround |
Restart MassHunter Workstation. |
|
SCR #9450 |
|
Title |
MassHunter Acquisition blacks out when adjusting Real-time Plot pane. |
|
Description |
On some computer systems, when you change the y-axis to its minimum scale using the scroll arrows, the screen becomes dark. |
|
Workaround |
Restart the PC. |
|
SCR #12696 |
|
Title |
Only 4 Scan Segments allowed when in MS2Scan Mode |
|
Description |
With the introduction of MassHunter Workstation B.01.04, 200 MRM Scan Segments are allowed. For scan or specifically using MS2Scan, there is a memory limitation constraining this to only 4 Scan Segments. This affects MS2Scan, Product Ion Scan, and Neutral Loss/Gain acquisition modes. |
|
Workaround |
Please restrict acquisition modes using MS2Scan to no more than 4 segments. |
|
SCR #12517 |
|
Title |
Acquisition Method sporadically fails to Import into MassHunter Workstation Quantitative Analysis. |
|
Description |
A batch created in MassHunter Workstation Quantitative Analysis with creating new Quant method from acquired MRM may produce the following error: "Failed to import method from acquisition Multiple method directories in <Userhome>\data..." There may be two acquisition method created. |
|
SCR #13225 |
|
Title |
Error may occur when starting offline Acq Method and Worklist Editor when acquisition is online. |
|
Description |
When MassHunter Workstation Acquisition is online, starting the Acq Method and Worklist may fail with "Error occurred when starting application or switching context. The application cannot continue and will abort”. |
|
SCR #13498 |
|
Title |
Solvent Manager fails if new solvent names have blank spaces in them. |
|
Description |
When using Solvent Manager, it has been found that naming conventions with blank spaces revert back to the default solvent setting. |
|
Workaround |
Do not use blank spaces for Solvent Manager naming convention. |
|
SCR #13687 |
|
Title |
Cannot Save Instrument Curves Settings at Next Startup. |
|
Description |
When acquiring, intrument curves can be selected that can include pump pressure, %A to %B, column temperature, etc. However, when MassHunter acqusition is started again, all these instrument curves selected are lost. Request a way to save the selected ones for the next analysis. |
|
SCR #13689 |
|
Title |
Ion Source and parameters are blank when offline Method for ESI with Agilent Jet Stream Technology is loaded for online acquisition. |
|
Description |
When a method is created in the offline Acq Method and Worklist editor for 6460 using ESI with Agilent Jet Stream Technology, ion source parameters are blank when used online for acquisition. |
|
Workaround |
Edit the ion source parameters for the 6460 using ESI with Agilent Jet Stream Technology. |
|
SCR #13694 |
|
Title |
Ghost Peak Sometimes Appear at 3000 m/z. |
|
Description |
There is a ghost peak in acquisition B.02.00 or B.02.01 pass the 3000 m/z. It is large and can tower over actual data in RTP. However, it seems to be only onscreen (and on RTP). One system starts to display this peak when in MS2 scan in acquisition from 50 to 1850. At 50 to 1800, all is normal. At 50 to 1850, the peak starts to take form. At 50 t0 1900, it becomes greater. At 50 to 3000, there is great abundance. |
|
SCR #13933 |
|
Title |
Only Contact 1 is visible in External Contacts tab for G1367D ALS SL+. |
|
Description |
When installing the contacts board on the G1367D ALS SL+ and launching MassHunter Workstation Acquisition, only Contact1 (out of 4) is visible. |
|
Workaround |
Change the existing method and save, and the External Contacts are available. |
|
SCR #13979 |
|
Title |
Needlewash location "flushseat" changes to "as method" when method is saved for G1367D ALS SL+. |
|
Description |
On G1367D ALS SL+ sampler, select "use injector program" in the Samples tab. Select the Injector Program tab, "needlewash" as the line function, "flushseat" from the location menu, and click apply and save the method. "Flushseat" changes to "as method" |
|
SCR #13981 |
|
Title |
MassHunter Workstation does not handle blank spaces. |
|
Description |
In relation to SCR#13498 with Solvent Manager, importing a worklist with blank spaces in the entry as created in Excel will fail to import. |
|
Workaround |
Do not use blank spaces for most MassHunter Workstation parameters. |
|
SCR #16146 |
|
Title |
Injector Program AND Overlapped Injection in Same Method Does Not Work. |
|
Description |
Injector program works fine alone. Start overlap at works fine alone. If both turned on, the overlapped injection never happens, and other strange things happen.
For example: First injection, both pumps (running ACR) never change to Run blue from the Waiting magenta color, MS runs stops at 0.7min rather than 2 min as specified in Pump1 tab. Next injections, all modules go Blue at start and run goes to 2min Stop, but no injection is ever made. |
3.4 Reports
|
SCR #13513 |
|
Title |
Wait for Autotune report to show before pressing OK when Autotune completes. |
|
Description |
Before the Autotune report is generated, pressing OK to the Autotune completed successfully window sometimes provides no report. |
|
Workaround |
Wait for Autotune report to show before pressing OK when Autotune completes. |
|
SCR #13832 |
|
Title |
MassHunter Workstation Acquisition Method report generation sometimes hang when printing. |
|
Description |
On new installations, MassHunter Workstation Acquisition Method report generation sometimes hang when printing the first time. This seems to be a more general issue with an Excel add-in. |
|
Workaround |
The Excel add-in loader for acquisition is shared with MassHunter Workstation Quantitative Analysis. Proceed to install any Quantitative Analysis software patch. |
Appendix A. Glossary
|
Term / Acronym |
Definition |
|
DAD |
Diode Array Detector |
|
QQQ |
Triple Quadrupole instrument. |
|
RF |
Radio Frequency |
|
RTP |
Real-time plot. This is the pane in the UI which includes the chromatogram and spectral plots. |
|
UI |
User Interface. The visual representation of the software with which a user interacts. |
|
VWD |
Variable Wavelength Detector |
|
|