Software Status Bulletin
G3793AA MassHunter Software rev. B.02.00
Revision Log:
|
Revision Date |
Reason For Update |
|
08/28/2008 |
Final release for B.01.04 revision (compatibility for MassHunter QQQ Acquisition B.01.04 Patch 1- 6410 only) |
|
11/11/2008 |
Final release for B.02.00 revision (compatible for MassHunter QQQ Acquisition B.02.00 for 6460 and B.02.01 for 6410/6460) |
1. Overview
The purpose of this document is to list known problems in the software and any known workarounds or recovery paths.
2. Known Problem Summary
The following is a summary of the known problems in the software. This is not an exhaustive list, but is a list of the problems a customer is most likely to encounter. This list cross-references items in our Defect tracking database.
|
SCR # |
Title |
|
12866 |
MassHunter Workstation Acquisition can be closed while MassHunter Optimizer is running. |
|
12876 |
“Optimization complete” message may be hidden. |
|
13201 |
Compound Browser Formula requires exact syntax and spaces to reference in table. |
|
13328 |
MassHunter Optimizer changes default method folder in MassHunter Acquisition. |
|
13679 |
“SaveAsProject” cannot overwrite as default. |
|
13703 |
“All records” progress bar does not work accurately. |
|
13840 |
Compound Setup Table does not maintain the order of each row when saved. |
|
13882 |
Project must be saved first to see the Summary Results View |
|
14119 |
G1367D ALS SL+ may have Invalid Plate Positions – B.02.00 |
|
14122 |
Some Project Information Settings are Lost when upgrading (B.02.00 from B.01.04) |
|
14187 |
Deleting Project Information may not be Complete. |
3. Known Problems Detailed Description
The following is the detailed descriptions of the known problems in the MassHunter Optimizer software. This section includes more extensive information from our defect tracking database. This section also includes any known workarounds or recovery paths
|
SCR #12866 |
|
Title |
MassHunter Workstation Acquisition can be closed while MassHunter Optimizer is running. |
|
Description |
One can inadvertently close MassHunter Workstation Acquisition while MassHunter Optimizer is running. There are no warnings or error messages, and Optimizer stops and needs to close. |
|
SCR 12876 |
|
Title |
“Optimization complete” message may be hidden. |
|
Description |
There is an “Optimization complete” pop up message that may be behind other open applications. In this case, the message is hidden, can be toggled to view, and then close. |
|
SCR 13201 |
|
Title |
Compound Browser Formula requires exact syntax and spaces to reference in table. |
|
Description |
In the Compound Browser, select Search Compounds/Formula and then enter a valid formula for a compound optimized. For example, use C66H95O26N1 which should reference the [Glu] fibrinopeptide b that had been optimized previously and now part of the database. Instead, spaces are required to correctly reference as C66 H95 O26 N19. How the compound was entered is how it will be referenced in the browser. |
|
Workaround |
Use exact syntax and spacing as entered to reference Compound Browser correctly |
|
SCR #13328 |
|
Title |
MassHunter Optimizer changes default method folder in MassHunter Acquisition. |
|
Description |
After running and closing MassHunter Optimizer, load a method in MassHunter. The default folder location is no longer D:\MassHunter\methods. |
|
Workaround |
Re-establish the method folder path by navigating to D:\MassHunter\methods. |
|
SCR #13679 |
|
Title |
“SaveAsProject” cannot overwrite as default. |
|
Description |
MassHunter Optimizer “SaveAsProject” cannot overwrite an existing project file name. |
|
Workaround |
“SaveAsProject” to a new project file name. |
|
SCR #13703 |
|
Title |
“All records” progress bar does not work accurately. |
|
Description |
While running MassHunter Optimizer, the “All records” progress bar (the second one on the bottom) does not update its status as expected. |
|
SCR #13840 |
|
Title |
Compound Setup Table does not maintain the order of each row when saved. |
|
Description |
Save a project after adding/importing some records with result in the Compound Setup tab (i.e.: Add CompoundRow, MethodRows, PrecursorIonRow, and ProductIonRows) and then close. Open the project again shows the order of each row of the Compound Table is not as it was when last saved. |
|
SCR #13882 |
|
Title |
Project must be saved first to see the Summary Results View. |
|
Description |
When the results are ready after optimization, the Summary Results View shows up empty (blank) unless project is saved first. |
|
Workaround |
Save project first to see Summary Results View. |
|
SCR #14119 |
|
Title |
G1367D ALS SL+ may have Invalid Plate Positions – B.02.00 |
|
Description |
When using the newer G1367D ALS SL+ sampler, the plate positions may be invalid. |
|
Workaround |
Downgrade the firmware on the G1367D LC module to 6.10 and configure as a G1367C temporarily. |
|
SCR #14122 |
|
Title |
Some Project Information Settings are Lost when upgrading (B.02.00 from B.01.04) |
|
Description |
When upgrading from B.01.04 to B.02.00 on 6410 Triple Quad LC/MS systems, some project information settings like all selected compound list, fine fragmentor, and etc. are lost. |
|
SCR #14187 |
|
Title |
Deleting Project Information may not be Complete. |
|
Description |
When deleting project information such as a product ion row and then saved, the information may still be visible after re-loading the project. This only occurs at first start up when the database is empty and for 1 project with 1 compound. |
Appendix A. Glossary
|
Term / Acronym |
Definition |
|
UI |
User Interface. The visual representation of the software with which a user interacts. |