Online literature is provided as Adobe Acrobat Portable Document Format (PDF) files. Obtain your free copy of the current Acrobat Reader from Adobe Systems.

Determination of Melamine Residue in Milk Powder and Egg Using Agilent SampliQ Polymer SCX Solid Phase Extraction and the Agilent 1200 Series HPLC/UV

Abstract: This method was developed for the determination of melamine in milk powder and egg. Solid phase extraction (SPE) and HPLC/UV used is consistent with the Chinese regulatory method. The sample preparation is performed using a polymeric mixed mode strong cation exchange resin. The separation and detection are performed by HPLC/UV. The limit of detection is 10 μg/kg. Linear calibration curves were obtained over a calibration range of 1 to 20 mg/kg. Overall recoveries range from 84 to 96 percent, with RSD values below 3.0 percent. ___________________________________________


Keywords: melamine, solid phase extraction, SPE,
Publication Number: 5990-3365EN
Last Updated: 1/7/2009
Number of Pages: 6