Online literature is provided as Adobe Acrobat Portable Document Format (PDF) files. Obtain your free copy of the current Acrobat Reader from Adobe Systems.

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN14105

Abstract: The analysis of free glycerin (glycerol) and total glycerides (mono-, di-, and triglycerides) in B100 biodiesel was performed according to ASTM method D6584 and CEN method EN14105. Method improvements were demonstrated through the use of a 530-μm id high-temperature fused-silica retention gap coupled to the analytical column. This was made possible with an Agilent Capillary Flow Technology Ultimate Union designed for inert, high-temperature GC oven operation. This configuration on the Agilent 7890A GC System showed calibration and precision performance that exceeded both D6584 and EN14105 specifications. This application provides complete system configuration as well as guidelines for successful analysis of free glycerin and total glycerides in biodiesel.


Keywords: Biodiesel, B100, ASTM D6584, CEN EN14105, ISO EN14105, glycerol, glycerin, monoglycerides, diglycerides, triglycerides, total glycerin, free glycerin, bound glycerin, FAME, fatty acid methyl esters,, Biodiesel
Publication Number: 5989-7269EN
Last Updated: 10/30/2009
Number of Pages: 8

For your convenience this PDF file will open in a separate window, and you may need to adjust your pop up blocker setting.