Specifications Agilent 1200 Series Variable Wavelength Detector SL Plus

Detection type

Double-beam photometer

Light source

Deuterium lamp

Noise

± 0.15 x 10 -5 AU, at 230 nm*

Drift

1 x 10-4AU/hr, at 230 nm

Linearity

>2.5 AU upper limit

Wavelength range

190-600 nm

Wavelength accuracy

± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter

Wavelength bandwidth

6.5 nm typical over whole wavelength range

Maximum data rate

160 Hz

GLP

Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.

Flow cells

Standard: 14 µL volume, 10 mm cell path length and 40 bar (588 psi) pressure maximum
High pressure: 14 µL volume, 10 mm cell path length and 400 bar (5880 psi) pressure maximum
Micro: 2 µL volume, 3 mm cell path length and 120 bar (1764 psi) pressure maximum
Semimicro: 5 µL volume, 6 mm cell path length and 40 bar (588 psi) pressure maximum

Communications

Controller-area network (CAN), RS-232. APG Remote: ready, start, stop and shut-down signals, LAN

Safety and maintenance

Extensive diagnostics, error detection and display through control module and Agilent ChemStation, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. Tracking of flow cells and lamps with RFID (radio frequency identification) tags

Housing

All materials recyclable


*Under specified conditions
 
 