|
Specifications 1200 Series Variable Wavelength Detector |
|
Detection type |
Double-beam photometer |
|
Light source |
Deuterium lamp |
|
Noise |
± 0.15 x 10 -5 AU, at 230 nm* |
|
Drift |
1 x 10-4AU/hr, at 230 nm |
|
Linearity |
>2.5 AU upper limit |
|
Wavelength range |
190-600 nm |
|
Wavelength accuracy |
± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter |
|
Wavelength bandwidth |
6.5 nm typical over whole wavelength range |
|
Maximum data rate |
20 Hz |
|
Flow cells |
Standard: 14-µL volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum High Pressure: 14-µL volume, 10-mm cell path length and 400 bar (5880 psi) Pressure maximum Micro: 2-µL volume, 3-mm cell path length and 120 bar (1764 psi) pressure maximum Semimicro: 5-µL volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum |
|
Communications |
Controller-area network (CAN), RS-232. APG Remote:ready, start, stop and shut-down signals, LAN |
|
Safety and maintenance |
Extensive diagnostics, error detection and display with Instant Pilot control module and Agilent ChemStation, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
|
GLP |
Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter. Tracking of flow cells and lamps with RFID (radio frequency indentification) tags. |
|
Housing |
All materials recyclable | *Under specified conditions
|
|