Agilent Technologies LSCA - Our measure is your success
clear
white
clear
white
clear

Specifications - 1200 Series Variable Wavelength Detector 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1


Specifications 1200 Series Variable Wavelength Detector

Detection type

Double-beam photometer

Light source

Deuterium lamp

Noise

± 0.15 x 10 -5 AU, at 230 nm*

Drift

1 x 10-4AU/hr, at 230 nm

Linearity

>2.5 AU upper limit

Wavelength range

190-600 nm

Wavelength accuracy

± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter

Wavelength bandwidth

6.5 nm typical over whole wavelength range

Maximum data rate

20 Hz

Flow cells

Standard: 14-µL volume, 10-mm cell path length and 40 bar
(588 psi) pressure maximum
High Pressure: 14-µL volume, 10-mm cell path length and 400 bar (5880 psi) Pressure maximum
Micro: 2-µL volume, 3-mm cell path length and 120 bar (1764 psi) pressure maximum
Semimicro: 5-µL volume, 6-mm cell path length and 40 bar
(588 psi) pressure maximum

Communications

Controller-area network (CAN), RS-232. APG Remote:ready, start, stop and shut-down signals, LAN

Safety and maintenance

Extensive diagnostics, error detection and display with Instant Pilot control module and Agilent ChemStation, leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.

GLP

Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter. Tracking of flow cells and lamps with RFID (radio frequency indentification) tags.

Housing

All materials recyclable


*Under specified conditions

 
 
grey1
related_bottom_left clear clear clear
clear
grey2