Performance Specifications 1200 Series standard autosampler
Valid when standard 100 µL metering head installed
Injection range 0.1- 100 µL in 0.1 µL increments
Up to 1500 µL with multiple draw (hardware modification required)
Precision <0.25% RSD from 5-100 µL, <1% from 1-5 µL, variable volume
Sample viscosity range 0.2 - 5 cp
Sample capacity 100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Injection cycle time Typically 50 s depending on draw speed and injection volume
Carry-over Typically < 0.1 %, < 0.05 % with external needle cleaning

Performance Specifications 1200 Series standard autosampler
Valid when standard 900 µL metering head installed
Injection range 0.1 - 900 µL in 0.1 µL increments (recommended 1 µL increments)
Up to 1800 µL with multiple draw (hardware modification required)
Precision Typically < 0.5 % RSD of peak areas from 5 - 2000 µL
Typically < 1 % RSD of peak areas from 2000 - 5000 µL
Typically < 3 % RSD of peak areas from 1 - 5 µL
Variable volume
Sample viscosity range 0.2 - 5 cp
Sample capacity 100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
Injection cycle time Typically 50 s depending on draw speed and injection volume
Carry-over Typically < 0.1 %, < 0.05 % with external needle cleaning
Pressure operating range 0-20 MPa (0-200 bar, 0-2950 psi)

 
 