Specifications Agilent 1200 Series Refractive Index Detector |
|
Detection type |
Deflection method |
|
Short-term noise |
±2.5 × 10-9 RIU |
|
Drift |
200 x 10-9 RIU/h |
|
Refractive index range |
1.00–1.75, calibrated |
|
Flow cell |
8 µL, 5 bar pressure maximum |
|
Temperature control |
Ambient +5 °C to 55 °C |
|
pH range |
2.3–9.5 |
|
Time programmable |
Polarity, peak width |
|
Zero adjustment |
Automatic zero |
|
Valves |
Automatic purge and automatic solvent recycle |
|
Analog output |
Recorder/integrator: 100 mV or 1 V, with offset adjustment, RIU range selectable |
|
GLP features |
Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits (purge interval) and feedback messages. Electronic records of maintenance and errors, automated OQ/PV procedures. |
|
|