Agilent 1200 Series Rapid Resolution LC System - Selected System Specifications
Pump delay volume
Configurable:
Low delay volume configuration: 120 µL
Medium delay volume configuration: 320 µL
Standard delay volume configuration: 600 - 800 µL
Flow rate range
0.05 - 5 mL/min
Maximum pressure
600 bar
Column dimensions
Length: 10 - 300 mm, ID: 0.05 - 8 mm
Temperature range
10 °C below ambient - 100 °C
Temperature stability
< ± 0.05 °C
Max. linear velocity
> 16 mm/sec (water / ACN gradient)
Injection volumes
0.1 - 40 µL (100µL and 1500 µL) without loop change, no overfill (drawn = injected volume)
Maintenance and system test
Front access to all maintenance parts
Maintenance instructions through multimedia CD-ROM
Time for full system test (OQ/PV) < 4 hours
Standard GLP features
Early maintenance feedback-EMF
(tracks lamp burn time, usage, number of injections with limits and feedback messages)
Extended GLP features (for systems inlcuding MWD SL, DAD SL or VWD SL Plus )
Data Recovery Card (DRC) for data security (prevents data losses in case of communication breakdowns)
Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions)
System control
Through local workstation software, client/server network data system and/or local handheld control module (1200 Series Instant Pilot). Default control: LAN / CAN.
Specifications of 1200 Series Rapid Resolution LC System Modules and Columns
Agilent ZORBAX Rapid Resolution HT Columns
Agilent 1200 Series micro vacuum degasser
Agilent 1200 Series binary pump SL
Agilent 1200 Series high-performance autosampler SL Plus
Agilent 1200 Series thermostatted column compartment SL
Agilent 1200 Series diode-array detector SL
Agilent 1200 Series multiple wavelength detector SL
Agilent 1200 Series variable wavelength detector SL Plus