Specifications Agilent 1200 Series Quaternary Pump
|
Flow |
|
Flow precision |
≤0.07 % RSD,or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature |
|
Flow accuracy |
± 1 % or 10 µL/min whatever is greater |
|
Range |
Setpoints from 0.001 to 10.0 ml/min, in 0.001 ml/min increments |
|
Pressure |
|
Operating range |
0 - 40 MPa (0 - 400 bar, 0 - 5880 psi) up to 5 ml/min 0 - 20 MPa (0 - 200 bar, 0 - 2950 psi) up to 10 ml/min |
|
Pulsation |
<2 % amplitude (typically <1 %) at 1 mL/min isopropanol at all pressures >1 MPa (> 10 bar, >147 psi) |
|
Compressibility compensation |
User selectable, based on mobile phase compressibility |
|
Gradient |
|
Delay volume |
800 - 1100 µL depending on back pressure |
|
Pulsation |
<0.20 % SD, at 0.2 and 1 mL/min |
|
Recommended pH range |
|
1.0 - 12.5 Solvents with pH <2.3 should not contain acids which attack steel |
|
GLP features |
|
Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. |
|