Specifications Agilent 1200 Series Preparative Pump

Flow precision

<0.3 % RSD in range from 1 mL/min to 100 mL/min

Hydraulic system

Dual pistons in parallel

Flow range

Setpoints from 0.001 to 100 mL/min, in 0.001 mL/min increments. Recommended: 5-100 mL/min

Pressure

Operating range 2–40 MPa (20–400 bar, 290-5880 psi) system pressure

Gradient formation

High pressure using two combined pumps

Delay volume

700 µL

pH range

1.0-12.5 recommended

 
 