Performance Specifications 1200 Series Micro Well-Plate Autosampler |
|
Injection range |
0.01 - 8 µL in 0.01 µL increments with the small loop capillary 0.01 - 40 µL in 0.01 µL increments with the extended loop capillary |
|
Precision |
Typically < 0.5 % RSD of peak areas from 5 - 40 µL, Typically < 1 % RSD from 1 - 5 µL Typically < 3 % RSD from 0.2 - 1 µL Variable volume |
|
Sample viscosity range |
0.2 - 5 cp |
|
Sample capacity |
2 well plates (96 and 384) plus 10 additional 2-mL vials 108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2-mL vials 30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2-mL vials 54 Eppendorf tubes (0.5/1.5/2.0mL) in 2 x 27 Eppendorf tube plate Also compatible with the Agilent 1200 Series sample capacity extension for further expansion of the sample capacity |
|
Injection cycle time |
Typically < 30 s using following standard conditions: Default draw speed: 4 µL/min Default eject speed: 10 µL/min Injection volume: 0.1 µL |
|
Carry-over |
Typically < 0.05 % using the following conditions: Column: 150 x 0.5 mm Hypersil ODS, 3 µm Mobile phase: Water/Acetonitrile = 85/15 Column Flow rate: 13 µL/min Injection volume: 1 µL caffeine (=25ng caffeine), 1 µL water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |
|
|