Agilent Technologies LSCA - Our measure is your success
clear
white
clear
white
clear

Specifications - 1200 Series Isocratic Pump 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1


Specifications Agilent 1200 Series Isocratic Pump

Flow

Precision

≤0.07 % RSD,or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow accuracy

± 1 % or 10 µL/min whatever is greater

Range

Setpoints from 0.001 to 10.0 mL/min,
in 0.001 mL/min increments

Pressure

Operating range

0 - 40 MPa (0 - 400 bar, 0 - 5880 psi) up to 5 mL/min
0 - 20 MPa (0 - 200 bar, 0 - 2950 psi) up to 10 mL/min

Pulsation

<2 % amplitude (typically <1 %) at 1 mL/min isopropanol at all pressures >1 MPa (> 10 bar, >147 psi)

Compressibility compensation

User selectable, based on mobile phase compressibility

Recommended pH range

 

1.0 - 12.5
Solvents with pH <2.3 should not contain acids which attack

GLP features

 

Electronic records of maintenance and errors.
Early maintenance feedback (EMF) for
continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded.

 
 
grey1