Specifications - 1200 GPC-SEC System 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1
 

 Agilent 1200 Series Pumps

Flow precision

< 0.1 % RSD or 0.02 min SD, whichever is greater, (typically < 0.065 % RSD or 0.015 min SD, six replicates, isocratic mode from 1 bottle, based on retention time

Pressure pulsation

< 2 % amplitude, (typically < 1 %) at 1 mL/min isopropanol, at all pressures > 1 MPa, (>10bar, 147 psi)

Compressibility
compensation

User-selectable, based on mobile phase compressibility

Recommended pH
range

1.0 – 12.5 pH range; Solvents with pH < 2.3 should not contain acids which attack stainless steel

Agilent 1200 Series Thermostatted Column Compartment

Temperature range

10 degrees below ambient to 80 °C

Temperature stability

± 0.15 °C

Temperature
accuracy

± 0.8 °C, with calibration ± 0.5 °C

Column capacity

Three 30-cm columns (if the application requires more columns, a second column compartment can be added)

Agilent 1200 Series Autosampler

Sample capacity

100 × 2 mL vials in 1 tray, 40 × 2 mL vials in 1/2 tray, 15 × 6 mL vials in 1/2 tray, Microvials (100 or 300 µL) with sleeves

Injection volume

0.1–100 µL injection range, up to 1500 µL with multiple draw
Optional injection range upgrade kit:
0.1–900 µL injection range in 0.1 µL increments (recommended 1 µL increments)
Up to 1800 µL with additional multiple-draw
Up to 5000 µL with additional exchange of seat capillary

Precision

Typically < 0.5% RSD from 5–100 µL, typically < 1% RSD from 1–5 µL,
Optional injection range upgrade kit
Typically < 0.5% RSD of peak areas from 10–900 µL

Agilent 1200 Series Refractive Index Detector

Detection type

Deflection method

Short-term noise

± 2.5 × 10-9 RIU*

Drift

200 × 10-9 RIU/h*

Valves

Automatic purge and automatic solvent recycle

* According to ASTM E-1303-95 "Practice for Refractive Index Detectors used in Liquid Chromatography." Reference conditions: response time 4 s, 35 °C, 1 mL/min H2O (on-line degassing), restriction capillary.

Specifications Common To All Modules

Communications

Controller-area network (CAN), RS-232, APG remote, optional local-area network (LAN). Remote: ready, start and shutdown signals

Safety and
maintenance

Extensive diagnostics, error detection and display, leak detection and safe leak handling, low voltages in major maintenance areas

GLP features

Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-settable limits and feedback messages. Electronic records of maintenance and errors, automated OQ/PV procedures.

 
 