Specifications 1200 Series Evaporative Light Scattering Detector |
|
Light source |
High efficiency Blue LED (470 nm) |
|
Temperature range (evaporator) |
Ambient to 100 °C |
|
Flow rate |
5 µL/min – 5 mL/min |
|
Nebulizers |
Useable flow ranges Large flow: 1 – 5 mL/min (useable flow rate) Standard flow: 0.2 – 2.0 mL/min (included) RRLC flow: 0.2 - 1.4 mL/min Semi-micro flow: 0.04 – 1.0 mL/min Micro-flow: 0.005 – 0.04 mL/min
High sensitivity and repeatability flow ranges Large flow: 2 – 3 mL/min Standard flow: 0.5 – 1.2 mL/min (included) RRLC flow: 0.4 - 1.0 mL/min Semi-micro flow: 0.1 – 0.3 mL/min Micro-flow: 0.015 – 0.025 mL/min |
|
Short-term noise |
< 2 mV |
|
Drift |
< 1 mV/h |
|
Signal output |
0 – 1 V (analog) RS 232 (digital) |
|
Gain settings |
211 (2048) |
|
Communications |
RS-232 Contact closure for ready, auto-zero APG remote start/stop |
|
Control and data evaluation |
Built-in local controller, Agilent ChemStation for LC (control and evaluation), EZChrom Elite (2008) |
|
Context sensitive help |
Agilent ChemStation for LC |
|
Gas flow control |
Manual and computer controlled (power down) nebulization gas flow and patented auxiliary gas flow |
|
Environment |
5 – 40° C; maximum humidity of 80 % for temperatures under 31° C, with linear decrease to 50 % at 40 °C |
|
Dimensions |
250 mm (10”) width x 450 mm (18”) height x 550 mm (22”) depth |
|
Weight |
18.5 kg (40 lb.) |
|
|