| Performance Specifications 1200 Series Dual Loop Autosampler |
| Injection modes |
Partial loop filling
Complete loop filling |
| Injection range |
0.1 - 5000 µL in 0.1 µL increments
Recommended injection range 10 - 5000 µL |
| Loop size |
Max. loop size 5 ml (selectable range: 5 µL - 50000 µL) |
Area precision
Complete loop fill |
Typically < 0.5% RSD,(50 µL loop, overfill factor 3 or greater) |
Area precision
Partial loop fill |
Typically < 1.0% RSD,(5000 µL loop, 1000µL injection volume,
draw/eject speed 20 mL/min) |
| Sample viscosity range |
0.2 - 5 cp |
| Sample capacity |
2 well-plates (96 and 384) plus 10 additional 2-mL vials
108 x 2-mL vials in 2 x 54 vial plate plus 10 additional 2-mL vials
30 x 6-mL vials in 2 x 15 vial plate plus 10 additional 2-mL vials
54 Eppendorf tubes (0.5/1.5/2.0 mL) in 2 x 27 Eppendorf tube plate
Vial septa and closing mats must be pre-slit. |
| Carry-over |
< 0.1% with rinse factor > 3 (see test method listing in manual) |
|
|
|
|