Specifications Agilent 1200 Series Diode-Array Detector SL

Detector type

1024-element diode array

Light source

Deuterium and tungsten lamp

Number of signals

8

Maximum sampling rate

80 Hz (both spectra and signals)

Short-term noise

< ± 0.8 x 10 -5 AU at 254 nm and at 750 nm

Drift

< 0.9 x 10 -3 AU/hr, at 254 nm

Linearity

>2 AU upper limit

Wavelength range

190-950 nm

Wavelength accuracy

± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter

Slit width

Programmable: 1, 2, 4, 8, 16 nm

Diode width

< 1 nm

Wavelength bunching

Programmable, 2 - 400 nm, in steps of 1 nm

Flow cells

Standard

13-µL volume, 10 mm cell path length, 120 bar (1760 psi) pressure maximum

Semi-micro

5-µL volume, 6 mm cell path length, 120 bar (1760 psi) pressure maximum

Micro

2-µL volume, 3 mm cell path length, 400 bar (5880 psi) pressure maximum

Semi-nano

500-nanoliter volume 10-mm cell path length, 50 bar (735 psi) pressure maximum

Nano

80 nanoliter volume., 6-mm cell path length, 50 bar (735 psi) pressure maximum

Preparative

3-mm cell path length, 120 bar (1760 psi) pressur maximum
0.3-mm cell path lenght, 20 bar (294 psi) pressure maximum, quartz
0.06-mm cell path length, 20 bar (294 psi) pressure maximum, quartz

High pressure

1.7-µL volume, 6-mm cell path length, 400 bar (5880 psi) pressure maximum

Time programmable

Wavelength, polarity, peak width, lamp bandwidth, autobalance,
WL range, threshold, spectra storage mode

Spectral tools

Data analysis software for spectra evaluation, including spectral libraries and peak purity functions

Analog output

Recorder/integrator 100 mV or 1 V, 2 outputs

Data recovery card to prevent data losses. RFID for electronics records of flow cell and UV lamp conditions (path length, volume, product number, serial number, test passed, usage). Electronic records of maintenance and errors. Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded. Verification of wavelength accuracy with built-in holmium oxide filter.

 
 