Agilent 1200 Series Capillary Pump
Hydraulic system Two dual pistons in series, with proprietary servo-controlled variable stroke drive, floating pistons and active inlet valve, solvent selection valve and electronic flow control (EFC) for flow rates up to 100 µL/min
Settable column flow range 0.01 – 20 µL/min, 0.01 – 100 µL/min (with extended flow range kit), 0.001 – 2.5 ml/min (electronic flow control bypassed)
Recommended column flow range 1–20 µL/min, 10–100 µL/min (with extended flow range kit)
Column flow precision < 0.7 % RSD or 0.03 SD, whatever is greater (typically < 0.4 % RSD or 0.02 SD), at 10 µL/min and 50 µL/min column flow (based on RT, default settings)
Composition precision < 0.2 % SD at 10 µL/min, 50 µL/min (with extended flow range kit) and 1 ml/min (default settings)
Pressure range up to 400 bar system pressure
Delay volume Typically 3 µL from EFC to pump outlet for flow rates up to 20 µL/min; 12 µL from EFC to pump outlet for flow rates up to 100 µL/min; 180–480 µL (system pressure dependant) without mixer for flow rates up to 2.5 ml/min; mixer delay volume 420 µL
pH range 1.0 to 8.5 recommended

 
 