Agilent 1200 Series Capillary LC System Specifications |
| Performance specs |
Baseline noise: ±3 × 10-5 AU, at 254 nm, at flow rates <100 µl/min |
| Maintenance and system test |
- Front access to all maintenance parts
- Maintenance instructions on a multimedia CD-ROM
- Time for full system test (OQ/PV) < 5 hours
|
| System delay volume (typical) |
5 µl from EFC to column head, for flow rates up to 20 µl/min (default setup); 14 µl from EFC to column head, for flow rates up to 100 µl/min (default setup) |
| Required benchspace |
< 36 cm |
| System control |
Through local computer software, LAN or local handheld control module Agilent 1200 Series instant pilot |
| GLP features |
Early maintenance feedback—EMF (tracks lamp burn time, usage, number of injections with limits and feedback messages) |
| Sample capacity |
100 × 2 ml vials in 1 tray, Microvials (100 or 300 µl) with sleeves |
Settable injection volume |
0.01–8 µl with small loop capillary; 0.01–40 µl with extended loop capillary |
| Precision |
Typically: < 0.5 % RSD from 5–40 µl, < 1 % RSD from 1–5 µl, < 3 % RSD from 0.2–1 µl |
| Carry-over |
Typically < 0.1 % without automated needle wash and 1 µl injection volume |
| Temperature range |
Settable from 4–40 degree C in 1 degree increments |
| pH-range |
1.0 to 8.5 recommended |
|
Agilent 1200 Series Modules |
|
|
|
Agilent 1100 Series LC/MSD family |
- Orthogonal-flow sprayer enhances sensitivity for a wide range of LC conditions and requires no adjustment
- Air-cooled vacuum system with dual turbo- molecular pumps – eliminates the need for cooling water
- All Agilent LC/MS detectors can be easily interfaced to capillary LC, capillary electro- phoresis and standard LC systems. Optional sprayers are available.
|
|
|