Specifications - 1200 Series Binary Pump 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1


Specifications Agilent 1200 Series Binary Pump

Flow

Flow precision

<0.07 % RSD,or ≤0.02 min SD whatever is greater, based on retention time at constant room temperature

Flow accuracy

± 1 % or 10 µL/min whatever is greater

Range

Setpoints from 0.001 to 5.0 ml/min, in 0.001 ml/min increments

Pressure

Operating range

0 - 40 MPa (0 - 400 bar, 0 - 5880 psi)

Pulsation

<2 % amplitude (typically <1 %) at 1 ml/min isopropanol at all pressures >1 MPa (> 10 bar, >147 psi)

Compressibility compensation

User selectable, based on mobile phase compressibility

Gradient

Delay volume

180 - 480 µl (600-900 µl with mixer) dependent on back pressure

Composition precision

<0.15 % RSD, 1 mL/min

Composition accuracy

<0.5 % absolute

Recommended pH range

 

1.0 - 12.5
Solvents with pH <2.3 should not contain acids which attack steel

GLP features

Electronic records of maintenance and errors.
Early maintenance feedback (EMF) for continuous tracking of instrument usage, display of feedback messages if preset limits are exceeded.

 
 