RNA 6000 Nano total RNA Kit |
RNA 6000 Nano mRNA Kit |
RNA 6000 Pico total RNA Kit |
RNA 6000 Pico mRNA Kit |
Small RNA Kit |
|
Analytical Specifications |
|
Quantitative range |
25–500 ng/µL |
25–250 ng/µL |
- |
- |
50-2000 pg/µL of purified miRNA in water |
|
Qualitative range |
5–500 ng/µL |
25–250 ng/µL |
50–5000 pg/µL in water |
250–5000 pg/µL in water |
50-2000 pg/µL of purified miRNA in water |
|
Sensitivity (S/N>3) |
5 ng/µL in water |
25 ng/µL in water |
50 pg/µL in water 200 pg/µL in TE |
250 pg/µL in water 500 pg/µL in TE |
50 pg/µL in water** |
|
Quantitation reproducibility |
10% CV (within a chip) |
10% CV (within a chip) |
20 % CV (within a chip) |
20 % CV (within a chip) |
25 % CV (within a chip) |
|
Quantitation accuracy |
20 % CV* |
20 % CV* |
30 % CV* |
30 % CV* |
- |
|
Maximum sample buffer strength |
100 mM Tris, 0.1 mM EDTA or 125 mM NaCl or 15 mM MgCl2 |
100 mM Tris, 0.1 mM EDTA or 125 mM NaCl or 15 mM MgCl2 |
50 mM Tris, 0.1 mM EDTA or 50 mM NaCl or 15 mM MgCl2 |
50 mM Tris, 0.1 mM EDTA or 50 mM NaCl or 15 mM MgCl2 |
10 mM Tris, 0.1 mM EDTA |
|
|
RNA 6000 Nano total RNA Kit |
RNA 6000 Nano mRNA Kit |
RNA 6000 Pico total RNA Kit |
RNA 6000 Pico mRNA Kit |
Small RNA Kit |
|
Physical Specifications |
|
Analysis time |
30 minutes |
30 minutes |
30 minutes |
30 minutes |
30 minutes |
|
Samples per chip |
12 |
12 |
11 |
11 |
11 |
|
Sample volume |
1 µL |
1 µL |
1 µL |
1 µL |
1 µL |
|
Kit stability |
≥4 months at 4 °C |
≥4 months at 4 °C |
≥4 months at 4 °C |
≥4 months at 4 °C |
≥4 months at 4 °C |
|
25 chips per kit |
RNA Nano 12/chip = 300 samples/kit |
RNA Pico 11/chip = 275 samples/kit |
- | * Determined analyzing the RNA ladder as sample ** Measeured for the 40 nt fragment of the Small RNA ladder
|