Agilent Technologies LSCA - Our measure is your success
clear
white
clear
white
clear

Specifications - 2100 Bioanalyzer Protein Kits 

white
clear clear clear clear clear

grey1
Literature Library
Technical Support
Education & Events

grey1
 

 

Protein 80 Kit

Protein 230 Kit 

High Sensitivity
Protein 250 Kit

Analytical specifications

Sizing range

5-80 kDa

14-230 kDa

10-250 kDa

Typical sizing resolution

10%

10%

10%

Typical sizing accuracy

10% CV (CAII, BLG)

10% CV (BSA, CAII)

10% CV (BSA)

Sizing reproducibility

3% CV (CAII, BLG)

3% CV (BSA, CAII)

3% CV (BSA)

Sensitivity
(Signal/Noise > 3)

6 ng/μL CAII (15 ng/μL BSA) in PBS, 10 ng/μL (CAII) in 0.5 M NaCl
(30 ng/μL BSA in 0.5 M NaCl)

6 ng/μL CAII (15 ng/μL BSA) in PBS
30 ng/μL (BSA) in 0.5 M NaCl

1 pg/μL (labeled BSA)
in water on chip
5 pg/μL (labeled BSA)
in PBS on chip
Labeling reaction at
1 ng/µL of total protein

Quantitative range

60-2000 ng/μL CAII in PBS

15-2000 ng/μL CAII, 30-2000 ng/μL BSA in PBS

up to 4 orders of magnitude (0.3 to 3000 ng/µL BSA)

Qualitative range

6-4000 ng/μL CAII and BLG

6-5000 ng/μL CAII, 15-5000 ng/μL BSA in PBS

-

Quantitation reproducibility

20% CV (CAII, BLG)

20% CV (BSA, CAII)

20% CV (BSA)

  

 

Protein 80 Kit

Protein 230 Kit

High Sensitivity
Protein 250 Kit

Physical specifications

Analysis run time

30 minutes

25 minutes

30 minutes

Number of samples

10 samples/chip

10 samples/chip

10 samples/chip

Sample volume

4 μL

4 μL

5 μL

Kit stability

4 months (for storage temperature see individual box)

4 months (for storage temperature see individual box)

6 months at -20° C

Compatible buffers

List of compatible buffers

List of compatible buffers

List of compatible buffers

CAII = Carbonic Anhydrase, BSA = Bovine Serum Albumin, BLG = beta-Lactoglobulin

 
 
grey1
related_bottom_left clear clear clear
clear
grey2